Flow injection determination of sodium, potassium, calcium, and magnesium in beer by flame emission and atomic absorption spectrometry

Flow injection systems for the determination of sodium and potassium, by flame emission photometry, and calcium and magnesium, by atomic absorption spectrometry, in beer are described. The samples were directly injected in the flow system where automatic dilution occurred, and reagent addition using...

ver descrição completa

Detalhes bibliográficos
Autor principal: Fernandes, Sílvia M. V. (author)
Outros Autores: Rangel, António O. S. S. (author)
Formato: article
Idioma:eng
Publicado em: 2011
Assuntos:
Texto completo:http://hdl.handle.net/10400.14/6584
País:Portugal
Oai:oai:repositorio.ucp.pt:10400.14/6584
Descrição
Resumo:Flow injection systems for the determination of sodium and potassium, by flame emission photometry, and calcium and magnesium, by atomic absorption spectrometry, in beer are described. The samples were directly injected in the flow system where automatic dilution occurred, and reagent addition using the merging zones technique was performed. Lanthanum(III) consumption was reduced in about 20-80-fold compared to conventional manual atomic absorption determinations. Sampling rates of 120-200 determinations/h were achieved. The results were in agreement with the reference procedures (relative deviations below 3.8%) and present good reproducibility (relative standard deviations lower than 1.6% for 10 consecutive measurements).