A gas proportional-scintillation counter for x-ray spectrometry in the 0.1-3 keV range

Results for the energy resolution of a uniform-field gas proportional-scintillation counter are presented for the range 0.1-3 keV. A comparison with results from the literature shows that the detector is most suitable for this energy range. Examples of x-ray fluorescence pulse-height distributions f...

Full description

Bibliographic Details
Main Author: Borges, F. I. G. M. (author)
Other Authors: Santos, J. M. F. dos (author), Dias, T. H. V. T. (author), Santos, F. P. (author), Conde, C. A. N. (author)
Format: article
Language:eng
Published: 2004
Online Access:http://hdl.handle.net/10316/8209
Country:Portugal
Oai:oai:estudogeral.sib.uc.pt:10316/8209