Comparison of two methods for coating piezoelectric crystals

Quartz piezoelectric crystals were coated with triethanolamine using two procedures: spray and syringe methods. The reproducibility of both coating methods was evaluated comparing the relative standard deviation of batches of four crystals coated by each of the procedures. The relative standard devi...

ver descrição completa

Detalhes bibliográficos
Autor principal: GOMES, MT (author)
Outros Autores: DUARTE, AC (author), OLIVEIRA, JP (author)
Formato: article
Idioma:eng
Publicado em: 1000
Assuntos:
Texto completo:http://hdl.handle.net/10773/6827
País:Portugal
Oai:oai:ria.ua.pt:10773/6827
Descrição
Resumo:Quartz piezoelectric crystals were coated with triethanolamine using two procedures: spray and syringe methods. The reproducibility of both coating methods was evaluated comparing the relative standard deviation of batches of four crystals coated by each of the procedures. The relative standard deviation obtained with the spray method is a hyperbolic function of the coating frequency and is, at least, three times lower than the one calculated from results of the syringe method, which is not frequency dependent. An explanation for the poorer reproducibility of the syringe method is given, based on the radial sensitivity theory and the non-uniformity of the coating distribution over the crystal, as evidenced by microscopic photographs.