On the assessment of time-shift variations from backscattered ultrasound for large temperature changes in biological phantoms

This work reports the assessment of time-shifts (TS) from backscattered ultrasound (BSU) signals when large temperature variations (up to 15 degrees C) were induced in a gel-based phantom. The results showed that during cooling temperature is linear with TS at a rate of approximately 74 ns/degrees C...

Full description

Bibliographic Details
Main Author: Teixeira, C. A. (author)
Other Authors: Ruano, M. Graça (author), Pereira, W. C. A. (author)
Format: conferenceObject
Language:eng
Published: 2015
Subjects:
Online Access:http://hdl.handle.net/10400.1/5863
Country:Portugal
Oai:oai:sapientia.ualg.pt:10400.1/5863
Description
Summary:This work reports the assessment of time-shifts (TS) from backscattered ultrasound (BSU) signals when large temperature variations (up to 15 degrees C) were induced in a gel-based phantom. The results showed that during cooling temperature is linear with TS at a rate of approximately 74 ns/degrees C. However during a complete heating/cooling cycle, the relation is highly non-linear. This can be explained by the fact that during cooling the temperature distribution is more uniform. Another problem to report is that TS is very sensitive to external movements.